APPLICATIONS:
- The Hg/HgO is the best choice reference for alkali or fluoride media
- Ideal for battery development
- Energy systems research and development
- Process control
- Electro-analytical studies
- Measuring half-cell potentials
- Titration, and many other electrochemical measurements.
STEPS TO VERIFY VIABILITY OF ELECTRODE:
- Create a ~5mm layer of electrode solution in a beaker
- Place two electrodes of the same type into the pool of solution
- Compare the readings using a volt meter
- If they are within 10 mV of variation, it is good to go!
FEATURES:
- Carry Case
- Electrode Refill Solution Electrolyte 30 ml
- Laboratory Electrode Storage Bottle
- Pack of 2 Replacement Porous Junctions
- Electrode Filling Pipette
- Absorbent Foam
- Maintenance Manual
- All plastic body parts available in two different sizes
- Choose either a banana jack or phone pin lead wire terminals
- Our standard concentration is 20% KOH (4.24 Molar) Potassium Hydroxide
- * (NaOH) Sodium Hydroxide electrolyte is also available
- Electrode potential is stable and recovers quickly from small temperature excursions.
- The porous junction tip is made with electroporous plastic- Low noise, low impedance, Prime ion transfer and is replaceable when damaged or contaminated.
- Traceable ID tags on the lead wire.
HOW IT WORKS:
The standard electrode potential of the Mercury/Mercury Oxide half-cell is +0.098 volts vs. the normal hydrogen electrode (NHE) in basic solution. The standard electrode potential of the Mercury/Mercury Oxide half-cell is +0.098 volts vs. the normal hydrogen electrode (NHE) in basic solution. The standard concentration of the electrode filling solution is 20% (4.24 Molar) Potassium Hydroxide (KOH). The electrode can be ordered in any concentration of Potassium Hydroxide not to exceed 30% KOH due to shipping regulations. Electrode potential is stable to plus or minus one millivolt and recovers quickly from small temperature excursions (about 125° C).
LEARN MORE:
- Electrode Only (No Kit)
- Reference Electrode Accessories
- Promotional Flyer
- Select the Right Koslow Reference Electrode
- What are Reference Electrodes?